Charles,
So if you have bound ligand with a covalently-linked Bromine, and the ligand isn't abundant in your unit cell, don't you *expect* the anomalous signal to be weak ? Have you tried a DANO model-phased difference map or something comparable ?
Phil Jeffrey
Princeton
On 7/10/14 12:52 PM, CPMAS Chen wrote:
Well, let me make this clear.
1. I am using Br anomalous signal to identify the potentially bound ligand.
2. I do shoot the crystals at 0.92A.
3. the different crystals have different resolution, but anomalous
signal was weak as reported by autoxds.
Charles
***************************************************
Charles Chen
Research Associate
University of Pittsburgh School of Medicine
Department of Anesthesiology
******************************************************